TSP #80 – Teardown, Repair & Experiments of an Agilent E7495B 2.7GHz Base Station Test Set

In this episode Shahriar repairs an Agilent E7495B 2.7GHz base-station test set capable of VSWR measurements as well as spectrum analysis. This particular unit is well equipped with many options however does not power on. A full teardown of the instrument reveals a faulty DC power conditioning module. The input filtering capacitor has failed. After component replacement, the unit is re-assembled. The functionality of the instrument is verified as well as a few quick experiments demonstrating various options.

Support TSP:
http://www.patreon.com/TheSignalPath

The Signal Path
http://www.TheSignalPath.com

Leave a Reply

Your email address will not be published. Required fields are marked *

This site uses Akismet to reduce spam. Learn how your comment data is processed.