TSP #42 - Teardown, Repair and Analysis of an Agilent E3642A DC Power Supply
In this episode Shahriar attempts a repair of an Agilent E3642A DC Power Supply which is completely non-responsive. After presenting a teardown of the power supply, the GPIB interface is used to verify the functionality of the power supply. The problem is traced to the main display unit which communicated with the main power supply via a serial interface.
After disassembly of the display, it is revealed that the entire unit has suffered a catastrophic failure due to the VFD display drive IC. All components must be individually removed and replaced. Unfortunately the main processor is a Mask ROM IC version (80C51) and cannot be sourced. Can you help Shahriar find a replacement part?