TSP #251 – Agilent 50GHz PSA (E4448A) Spectrum Analyzer Teardown, Repair & Experiments (Part 1)

In this episode Shahriar takes a look at the faulty Agilent 50GHz spectrum analyzer which does not power on. After tracing the issue and restoring the bootting sequence, it is clear that the instrument has several other issues. The attenuator step calibration failure is addressed which clears all startup errors. However, the instrument still does not measure some frequency bands correctly which points to a failure in the band-selection & mixer assembly. We will have to return to this instrument again in Part 2!