TNP #32 - Fluke 732A DC (10V, 1.018V, 1V) Reference Standard Quick Repair
In this episode Shahriar quickly repairs a faulty Fluke 732A DC Reference Standard. Although the unit does produce a good 10V and 1.018V output, the 1V output shows no voltage at all. A close examination of the schematic shows that the 1V output is derived through a resistor divider from the 10V source.
Since no resistance can be measured from the 1V to any of the other ports, it is suspected that the failure is caused by a broken connection behind the front panel. The connection is corrected and the unit produces all desired output signals.