TSP #321 – Lab Upkeep Time! Agilent 3458A 8.5 Digit Multimeter & MSO9404A 4GHz, 20GS/s Scope Repair

In this quick episode Shahriar performs some regular lab maintenance. Two instrument need attention. The Agilent 3458A is reporting a Nonvolatile Memory Error & and the Agilent MSO9404A is reporting a dead CMOS battery.

Both problems are easily addressed by changing the battery backup. In the case of the 3458A, the calibration content has to be re-programed into the new NVRAM ICs.

Please consider supporting The Signal Path:
https://www.TheSignalPath.com/Donate
https://www.Patreon.com/TheSignalPath

Other links:
https://www.TheSignalPath.com
https://www.Twitter.com/TheSignalPath

TSP #320 – Agilent 6GHz MXG Vector Signal Generator (N5182A) Teardown, Repair & Frac-N PLL Analysis

In this episode Shahriar repairs an Agilent N5182A MXG Vector Signal Generator. The instrument produces the correct output signal for some set frequencies but reports a Fractional-N Unlock error for others while the output signal completely disappears. The instrument also fails many self-test checks.

The block diagram of the unit is described in details with a measurement strategy to analyze every element of the PLL loop. The band-pass filter banks are isolated and measured with a Siglent network analyzer and are shown to be fully functional. The VCO output amplifiers & doubler are also fully functional and tested using a Siglent spectrum analyzer. The PLL loop is then "overwritten" and manually adjusting the VCO control voltage demonstrates that the oscillator can cover its entire designed tuning range. The fractional-N divider also appears functional for the appropriate input frequency ranges. Finally, although the GaAs phase detector appear functional, it is possible for such a component to slowly degrade where the phase detection gain changes significantly. This custom Agilent part is sourced from a donor board and the instrument now functions across all frequency ranges and passes all self-tests.

The phase noise performance of the unit is verified against the datasheet using an Agilent E5052B signal analyzer with cross-correlation capability.