Archive for December 24, 2022

TSP #245 – Magic of Mechanical Timekeeping & Measuring the Accuracy of a Raymond-Weil Skeleton Watch

In this episode Shahriar takes a look at the Raymond-Weil Blue Skeleton Freelancer wristwatch. The watch runs a little too fast. But before measuring its accuracy the internal structure of the watch as well as a few unique imaging of the watch mechanism are presented.
The teardown, theory of operation & measurements using a Multifunction Time Grapher No. 1000 is also presented. By using an external speaker in the place of the watch, the 8GHz oscillation of the watch can be emulated. It appears that the wristwatch runs 1mHz too fast resulting in gaining of about 12 seconds per day.

Make sure you check out @WristwatchRevival also!

http://www.TheSignalPath.com
http://www.YouTube.com/TheSignalPath​​​
http://www.Patreon.com/TheSignalPath

TSP #244 – Agilent 81101A 50MHz 20Vpp Pulse Generator Teardown & Repair

In this episode Shahriar takes a close look at an Agilent pulse generator which does not produce any output signal. The generator has an output ON/OFF function which can disconnect the unit using a microwave relay. The output spectrum of the faulty unit shows the presence of the desired spectrum and residue of the square wave signal.
The teardown shows that the output relay is not being engaged with the output is enabled. The fault is traced to the relay switch transistor (NPN) which does not respond to the control signal. The replacement NPN corrects the problem and the unit produces the correct output. The unit's maximum output swing as well as frequency accuracy are verified using a R&S MXO-4 Series & a Rubidium frequency counter.

http://www.TheSignalPath.com
http://www.YouTube.com/TheSignalPath​​​
http://www.Patreon.com/TheSignalPath

TSP #243 – Keysight E36313A Triple Output Programmable DC Power Supply Teardown & Repair

In this episode Shahriar repairs a broken Keysight E36313A with a blown fuse and burnt smell coming from the unit. The teardown shows the dual-board design of the unit as well as the large toroidal transformer. The magic smoke has escaped from a dual Schottky diode power device with a Vf = 0.45V at If = 5A. The replacement eBay component is first verified to meet the specifications and installed in the power supply. The Keysight unit is then tested at a few output voltages and current to ensure proper functionality.

http://www.TheSignalPath.com
http://www.YouTube.com/TheSignalPath​​​
http://www.Patreon.com/TheSignalPath

TSP #242 – JouleScope JS220 Precision Energy Analyzer Scope Review, Teardown & Experiments

In this episode Shahriar takes a close look at the new JouleScope JS220 dynamic energy analyzer:

http://joulescope.com

The USB-C instrument architecture and details teardown is presented alongside the main PCB X-Ray images. The instrument's unique six orders of magnitude current measurement and switching networks is also described and analyzed. The measurements confirm the six unique shunt resistors (alongside pairing CMOS devices) which can be chosen both manually and automatically by the autorunning software.
The GUI multimeter view is used to verify the accuracy of the unit across all current ranges and voltage ranges using a pair of Keithley 2450 SMUs. The dynamic behavior of the instrument is presented using a wide dynamic range load using various excitation waveforms and frequencies. Finally, a Z-Wave IoT motion sensor characteristics and communication patters is analyzed.

http://www.TheSignalPath.com
http://www.YouTube.com/TheSignalPath​​​
http://www.Patreon.com/TheSignalPath