In this episode Shahriar repairs a faulty Agilent 3458A. These highly desirable multimeters have continued to climb in price in the past few years. The last time The Signal Path has repaired one of these units was back in 2018.
This instrument generates several errors both during power up as well as Auto Cal procedure. The error points to a failure in the OHMS measurement mode. Detailed instrument block diagram and schematics are examined. Measurements also show incorrect DC current sourcing during OHMS measurements and the associated circuits are closely analyzed. The problem is traced to a quad-comparator IC. Replacement corrects all problems during ACAL. The battery backup of the SRAM ICs are also replaced to eliminate non-volatile errors. The IC of the comparator chip is also extracted and examined under the microscope.
The repaired instrument performance is verified by using voltage, current and resistance standards.
In this episode Shahriar repairs an Agilent N9912A which flags several error messages after bootup. The instrument reports hardware time-outs and occasionally ADC clock source unlock condition. The second error is much more useful in debugging the unit.
The block diagram of the instrument is presented and several potential failures anticipated.
The teardown of the unit and analysis is also presented. After some troubleshooting, it becomes clear that the primary 30MHz TCXO has a broken internal connection. The unit is disassembled and new pins installed. After putting the TCXO back into the unit, the instrument return to full functionality. A few quick experiments verify that the FieldFox performs both CAT and SA modes correctly.
In this episode Shahriar shows an extensive review of the Sanko (BirdRF) GH60 GeneHawk handheld vector signal generator which operates from 10MHz – 6GHz. With up to 100MHz of modulation bandwidth (licensed) the unit has good potential for a portable wide-band communication signal generator:
In this episode Shahriar takes a look at a faulty dry block calibrator from Ametek. These Jofra series temperature calibrators are widely used for thermostat validation as well as material analysis. This particular unit does not cool down or warm up but seems otherwise functional.
A brief set of measurements shows that one of the Peltier elements is damaged and should be replaced. A replacement is located and the temperature well is re-sealed. The unit is functional and the minimum temperature performance is verified.
The Signal Path (TSP) is a science & electrical engineering video blog for industry professionals, researchers & students. TSP is a not-for-profit website dedicated to provide free education spanning a wide range of electrical engineering & science topics. Equipment reviews, tutorials and repair videos are posted regularly.