In this episode Shahriar repairs an Agilent E5071C ENA Vector Network Analyzer generously loaned by AllTest. Please visit their website for all your measurement, calibration and service needs:
sales@alltest.net
https://alltestchambers.net
https://alltest.net
The network analyzer is missing the main HDD which initially prevents it from being tested. An image from a donor E5701C instrument is used to boot the unit and gain access to the instrument firmware. The unit shows an inaccurate S11 with an open Port 1 while S22 shows a perfect reflection. This indicates a potential problem with the Port 1 T/R module.
Teardown of the module reveals catastrophic damage to the resistive front-end coupler which has been mostly likely caused by a large RF input power incident. After cleaning the PCB area, a replacement coupler is constructed. The repaired unit is checked for full self-test and used to measure a YIG tuneable filter.
You can may check the Andostar microscope lineup here: (Direct Link / Not-tracked)
https://www.aliexpress.com/store/900802?spm=a2g0o.detail.100005.2.4d4f407ekYxi5H
More information about Wheatstone Bridge:
https://www.mwrf.com/technologies/test-measurement/article/21849280/the-wheatstone-bridge-how-does-it-impact-vna-measurements
http://www.TheSignalPath.com
http://www.YouTube.com/TheSignalPath
http://www.Patreon.com/TheSignalPath









Leave a Reply